Kent Imaging, Inc.: Autofluorescence Imaging Adjunct Tool For Wounds

Recall Summary Kent Imaging's SnapshotGLO Wound Imaging Device (Models KB100 and KB100AA) is being recalled because a software problem prevents the device from charging after its battery fully drains into a low-power shipping mode. When this happens, the device stops working for wound imaging until the issue is fixed. This Class II recall affects devices distributed across the United States.

Recalling firm
KENT IMAGING, INC.
Product
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
Reason
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-30
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/ccc21bbc-4730-4d18-b0e9-ec7959edab94