Philips Medical Systems Nederland B.v. Veenpluis: Interventional Fluoroscopic X-ray System
Recall Summary Philips is recalling several models of Integris-Allura interventional X-ray systems due to a deaeration hose in the X-ray tube cooling unit that may degrade and leak oil over time. When oil leaks, the cooling system's performance can drop, causing the system to automatically switch to low-dose fluoroscopy mode with an error message that cannot be fixed by restarting. This recall affects systems manufactured between February 2016 and May 2020 and distributed nationwide in the U.S. and internationally.
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis
- Product
- Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044), Integris-Allura 9 (722018), Integris-Allura 9 (Biplane) (722021), Integris CV Cesar-Powerpack (722030), Integris H5000 C / Allura 9C (722016);
- Reason
- The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-04-28
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican...