Dxc 500i Au Clinical Analyzer
- Recalling firm
- Beckman Coulter NZ Ltd
- Product
- DxC 500i AU Clinical Analyzer. [Device]. Model: With ISE - AU w/ISE and Without ISE -AU w/out ISE - software 1.5.0 and below
- Reason
- The manufacturer has identified an issue that may affect the reporting of Clinical Chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a Clinical Chemistry test and multiple tests are ordered under the same Sample ID, or multiple replicates of the same test under the same Sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-11-10
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory