Dxc 500i Au Clinical Analyzer

Recalling firm
Beckman Coulter NZ Ltd
Product
DxC 500i AU Clinical Analyzer. [Device]. Model: With ISE - AU w/ISE and Without ISE -AU w/out ISE - software 1.5.0 and below
Reason
The manufacturer has identified an issue that may affect the reporting of Clinical Chemistry test results. When a laboratory selects a reporting unit other than the system regional default unit for a Clinical Chemistry test and multiple tests are ordered under the same Sample ID, or multiple replicates of the same test under the same Sample ID, the software may incorrectly apply the unit conversion factor more than once. This may result in an erroneous reported value.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-10
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/cbc1a945-85a7-4841-8f61-7607f77ca8e8