Accuray Incorporated: Accelerator, Linear, Medical

Recall Summary Accuray is recalling certain Cyberknife radiation treatment systems and linear accelerators worldwide because a collimator (a component that shapes radiation beams) may fail to fully dock properly. When this happens, the system's safety lock prevents operation from the main console, but if someone uses the in-room control pendant to start motion anyway, the collimator can detach and fall from the equipment. This recall affects medical facilities across the US and numerous countries internationally.

Recalling firm
Accuray Incorporated
Product
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Reason
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-16
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/cb9e03ae-0fb4-4bc5-82b3-928e52cf9352