Infant Resuscitation System Correction: Ge Healthcare Issues Correction For Giraffe And Panda Infant Resuscitation Systems And Warmers With A M1091607-r Blender

Recall Summary GE HealthCare is correcting certain Giraffe and Panda Infant Resuscitation Systems and Warmers equipped with an M1091607-R Blender because the air-oxygen blender knob shaft may loosen over time. This loosening can cause the oxygen concentration delivered to patients to become inaccurate. The correction affects healthcare facilities that use these specific infant resuscitation systems.

Product
Giraffe and Panda Infant Resuscitation Systems and Warmers with a M1091607-R Blender
Reason
The air-oxygen blender knob shaft on certain integrated and stand-alone Resuscitation systems can loosen, affecting the delivered oxygen concentration
Category
drug recall
Country
United States
Agency
FDA-MEDWATCH-RSS
Recall date
2026-06-22
Status
Ongoing

Official recall notice from FDA-MEDWATCH-RSS

Recall details on Nibble Recalls: https://trynibble.app/recalls/caf806c4-783b-422c-b190-01cd6ceb9948