Fresenius Medical Care Holdings, Inc.: Dialysate Concentrate For Hemodialysis (liquid Or Powder)

Recall Summary Fresenius Medical Care is recalling Bicarby dialysate concentrate (used in hemodialysis treatment) in six specific model numbers because the luer-lock connections are leaking when the product is being prepared for use. The leaking fluid can create a slip and fall hazard on floors. This recall affects dialysis patients and facilities across the United States.

Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;
Reason
Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-12
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of NY, TN and TX.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/ca390820-d81d-46db-b9d5-3422bc62e848