Recall Summary UCB Japan is recalling multiple batches of Simzia 200mg subcutaneous injection (an auto-injector pen used to treat rheumatoid arthritis and certain skin conditions) because the needle may not fully retract after injection in some units. The affected batches were distributed between September 2025 and May 2026, with expiration dates ranging from October 2026 to June 2027. While a needle stick injury is possible if the exposed needle is contacted after use, the company states no serious health effects have been reported and the risk of severe infection is considered very low.
- Product
- ユーシービージャパン「シムジア皮下注200mgオートクリックス」 - 回収
- Classification
- Unclassified
- Category
- consumer product recall
- Country
- Japan
- Agency
- CAA
- Recall date
- 2026-10-15
- Status
- Ongoing
- Distribution
- Japan