Ferno-washington Inc: Stretcher, Wheeled

Recalling firm
Ferno-Washington Inc
Product
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091
Reason
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-19
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Netherlands, Canada, Germany, Norway, United Kingdom, Denmark and New Zealand.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/c8ffb994-95a7-4a5d-846a-115519bf6bcd