MedChart Recall Summary MedChart software used in New Zealand hospitals, pharmacies, and laboratories is being recalled due to a required software upgrade. This is classified as a Class I recall, the most serious category. The recall is being managed by Dedalus (Australia) Pty Ltd.

Recalling firm
DH (New Zealand) Limited
Product
MedChart. [Device]. Affected: MedChart v12.6.1
Reason
The manufacturer has identified that when calculating a dose from a QuickList/Protocol with the weight selected as Ideal Body Weight (IBW) or Adjusted Body Weight (AdjBW) for an adult (male or female), the dose is being calculated using the corresponding Child IBW or AdjBW weight formula instead of the adult calculator, resulting in a lower-than-expected dose.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-05-08
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE