Munevo Gmbh: Components, Wheelchair

Recall Summary MUNEVO GMBH is recalling munevo DRIVE and munevo DRIVE r-net powered wheelchairs with hands-free control systems due to a software problem that can cause the wheelchair to lose steering calibration. When this happens, the wheelchair may steer unexpectedly or not respond precisely to commands, potentially causing it to veer off course and result in collisions, tipping, or injury. The affected wheelchairs were distributed nationwide in the U.S. and in the United Kingdom, Spain, Italy, Germany, and Switzerland.

Recalling firm
MUNEVO GMBH
Product
munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System
Reason
A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-19
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.