Allura Angiographic X-ray

Recalling firm
Philips New Zealand Commercial Limited
Product
Allura Angiographic X-Ray. [Device]. Model: System code: 722003 - Allura Xper FD10; System code: 722079 - Azurion 7 M20, System code: 722221 - Azurion 3 M12, System code: 722224 - Azurion 7 M20. Also covers: [MDR] Azurion
Reason
The manufacturer has identified through customer complaints that, during certain repair or service actions, HDDs with different performance characteristics than those specified by the legal manufacturer may have been installed. These HDDs may exhibit reduced reliability over time, which may result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorised movement or loss of data., In addition, the manufacturer has identified that in some Allura systems, HDDs compliant with specifications but of different types have been used within a single...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-07-02
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/c63eb856-f258-4639-9bc4-e4ee40c511bf