Medline Convenience Kits: 1) Chest/breast-lf, Model Number: Dynj83950

Medline Convenience Kits Recall Summary Medline is recalling Chest/Breast-LF Convenience Kits (Model DYNJ83950) because calibration problems with the sterilization and packaging equipment may have affected the sterility of these products. While the kits went through sterilization and packaging processes, the equipment calibration issues mean they may not meet required sterility standards. This recall affects all units of this model that were processed with the affected equipment.

Product
Medline Convenience Kits: 1) CHEST/BREAST-LF, Model Number: DYNJ83950
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class II
Category
consumer product recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-01-07
Status
Open, Classified

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/c610809c-6ed9-493f-9ed1-ef96b4518e4f