General Purpose Dental Tray

Product
Medline Convenience Kits: 1) DYNDA3003, Model Number: DENTAL PACK; 2) DYNJ87157, Model Number: GULLO SURGICAL PACK
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-01-07
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/c53517e7-ec0a-41e8-9719-6a70c5e778e2