Ge Medical Systems, Llc: Incubator, Neonatal

Recall Summary GE HealthCare is recalling Giraffe OmniBed Carestations (Model 2082844-001-XXX), which are devices that combine an infant incubator and warmer used in hospitals. Fasteners securing the canopy soffit under the control panel can come loose or fall into the patient area during transport or impact, potentially injuring newborns. The recall affects devices distributed nationwide across all U.S. states and territories.

Recalling firm
GE Medical Systems, LLC
Product
Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubator and an infant warmer, Model Numbers 2082844-001-XXX
Reason
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-01
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/c45b0034-45f0-45c6-946d-8682261e17c2