Intuitive Surgical - Robotic Forceps
- Recalling firm
- Device Technologies New Zealand
- Product
- Intuitive Surgical - Robotic forceps. [Device]. Model: Affected Product: 470171, 470344, 471171, 471344, 471405; Affected Product: 470401, 470230, 470327; Affected Product: 470194, 470309, 471296, 471309; Affected Product: 470001, 470007; Affected Product: 470246, 470249. Affected: Affected Product: 470033, 470036, 470048, 470093, 470181, 470190, 470347, 471048, 471190. Also covers: Intuitive Surgical - Bipolar instruments; Intuitive Surgical - Clip applier; Intuitive Surgical - Needle driver; Intuitive Surgical - Robotic scissors; Intuitive Surgical - Retractor
- Reason
- The grip cables that control the opening and closing of the jaws of the instrument can fail partially (i.e. frayed) or completely (i.e. broken). A broken grip cable can lead to loss of grip functionality, exposure to frayed cables, or the potential for tungsten cable particulate to fall into the patient. If a cable were to fail, it would be retained within the shaft of the instrument. As a result, fragments would not fall into the patient, though particulate may be generated. A partial failure might not affect grip functionality but may lead to exposure to frayed cables., As part of a...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-15
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE