Abre Venous Self-expanding Stent System
- Recalling firm
- Medtronic New Zealand Limited
- Product
- Abre Venous Self-Expanding Stent System. [Device]. Model: All Models. Affected: All Lots
- Reason
- The manufacturer has received reports of partial stent deployments. Most events resulted in no or minimal patient harm, although some required additional intervention, including endovascular procedures, surgical conversion, or extended medical treatment., An update to the Instructions for Use (IFU) for the Abre⢠Venous Self-Expanding Stent System has been issued. This update provides a validated manual deployment workaround method to help mitigate potential harms in the rare event they experience a partial stent deployment.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-13
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE