Abre Venous Self-expanding Stent System

Recalling firm
Medtronic New Zealand Limited
Product
Abre Venous Self-Expanding Stent System. [Device]. Model: All Models. Affected: All Lots
Reason
The manufacturer has received reports of partial stent deployments. Most events resulted in no or minimal patient harm, although some required additional intervention, including endovascular procedures, surgical conversion, or extended medical treatment., An update to the Instructions for Use (IFU) for the Abreā„¢ Venous Self-Expanding Stent System has been issued. This update provides a validated manual deployment workaround method to help mitigate potential harms in the rare event they experience a partial stent deployment.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-10-13
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/c1b66ecd-6d78-4dee-99d5-3d7ecc43dbbe