Steris Corporation: Rapicide Pa Part A High-level Disinfectant, Model/catalog Number: Ml020117
Recall Summary Steris Corporation is recalling RAPICIDE PA Part A High-Level Disinfectant bottles (Model ML020117) because a blocked vent cap can cause gas to build up inside the bottle, making it swell. When the cap is opened, this pressure may release vapor or liquid that could expose users to the disinfectant chemical if they're not wearing proper protective equipment. This product was distributed nationwide across the United States.
- Recalling firm
- Steris Corporation
- Product
- RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117
- Reason
- If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-09-21
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide distribution.