Baxter Healthcare Corporation: Device, Positive Pressure Breathing, Intermittent

Recall Summary Baxter is correcting Volera Positive Pressure Breathing Devices because patient circuits may leak air and medication from the nebulizer cup during use. This leakage can reduce oxygen delivery and prevent medication from being delivered effectively to patients. The correction affects users of Volera breathing devices.

Recalling firm
Baxter Healthcare Corporation
Product
Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No
Reason
Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-08-03
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of Colorado and Florida.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/bf72f0ab-fe80-46c6-b72f-520292136c22