Beckman Coulter Inc.: System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Recall Summary Beckman Coulter is recalling several MicroScan laboratory analyzer models because they contain incorrect algorithms that may produce inaccurate antimicrobial susceptibility test results, potentially showing bacteria as susceptible to antibiotics when they may not be. The recall affects five analyzer models distributed nationwide across the United States. This is classified as a Class II recall, meaning the problem could result in serious health consequences but is unlikely to cause death.

Recalling firm
Beckman Coulter Inc.
Product
REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument
Reason
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-26
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.

Official recall notice from FDA-DEVICES