Medline Industries, Lp: Eye Tray

Recall Summary Medline is recalling eight medical procedure kits used in eye surgery (retina, vitrectomy, and tonsil procedures) because manufacturing samples failed to meet performance standards, including surface finish irregularities, premature connector detachment, and incorrect catheter diameter in some cases. The affected kits were distributed nationwide in the US and to several countries including Canada, the Netherlands, and others. This is a Class II recall, meaning there is a potential health risk but not an immediate life-threatening situation.

Recalling firm
Medline Industries, LP
Product
MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775K; 4) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775L; 5) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775M; 6) TONSIL PACK, Medline Kit Number/SKU DYNJ84480; 7) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080; 8) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080A.
Reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-04
Status
Open, Classified