Pride Mobility Products Corp.: Wheelchair, Powered
Recall Summary Quantum Rehab is recalling Quantum 4Front 2 powered wheelchairs with software version 2.0.11 because reminder pop-up messages can unexpectedly appear while the chair is in use, potentially causing the wheelchair to suddenly stop or its control system to crash. This could result in falls or injuries to users. The recall affects wheelchairs distributed nationwide in the US, as well as in Canada and Australia.
- Recalling firm
- Pride Mobility Products Corp.
- Product
- Brand Name: Quantum Rehab Product Name: Quantum 4Front 2 Model/Catalog Number: Quantum 4Front 2 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
- Reason
- If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-06-25
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and...