Cooper - Apple Hunts
- Recalling firm
- Obex Medical Ltd
- Product
- Cooper - Apple Hunts. [Device]. Model: 5MM APPLE HUNT SECONDARY CANNULA PYRAMIDAL TROCAR. Affected: Product code: CO-900-800 - Affected lot number: 550034148
- Reason
- The manufacturer is recalling a specific lot of the Apple-Hunt Secondary Cannula™ Pyramidal Tip (5mm) following reports of devices that may contain a cannula component that is longer than the intended specification. As a result, when the device is assembled in accordance with the Instructions for Use, the pyramidal trocar tip does not extend beyond the distal end of the cannula as intended., The condition is readily detectable during the required assembly and pre-use inspection process, as the trocar tip remains enclosed within the cannula following assembly.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-13
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/bcfe4e47-66e0-4af2-a85d-8684ddfda86b