Product
Cooper - Apple Hunts. [Device]. Model: 5MM APPLE HUNT SECONDARY CANNULA PYRAMIDAL TROCAR. Affected: Product code: CO-900-800 - Affected lot number: 550034148
Reason
The manufacturer is recalling a specific lot of the Apple-Hunt Secondary Cannula™ Pyramidal Tip (5mm) following reports of devices that may contain a cannula component that is longer than the intended specification. As a result, when the device is assembled in accordance with the Instructions for Use, the pyramidal trocar tip does not extend beyond the distal end of the cannula as intended., The condition is readily detectable during the required assembly and pre-use inspection process, as the trocar tip remains enclosed within the cannula following assembly.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-10-13
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory