Generic Name: Mobile Digital General-purpose Integrated X-ray Fluoroscopy System — Trade Names: (1) Corevision 3d X-ray Fluoroscopy System — (2) Corevision Ld X-ray Fluoroscopy System — (3) Corevision Sd X-ray Fluoroscopy System

Recalling firm
Name of Manufacturer and Distributor: Fujifilm Corporation — Address of Manufacturer and Distributor: 798 Miyadai, Kaisei-machi, Ashigarakami-gun, Kanagawa Prefecture — Type of License: Class I...
Product
Lot: (1) X線透視診断装置 CoreVision 3D 製造番号: 28000, 28002, 28014, 28018, 28019, 28021, 28023, 28024, 28026, 28029, 28031, 28032, 28033, 28034, 28035, 28037, 28039, 28041, 28042, 28045, 28046, 28048, 28049, 28051, 28060, 28064, 28068, 28069, 28073, 28074, 28075, 28078, 28079, 28080, 28081, 28082, 28087, 28089, 28090, 28092, 28094, 28096, 28103, 28107, 28108, 28109, 28110, 28111, 28112, 28113, 28114, 28118, 28121, 28123, 28125, 28126, 28127, 28128, 28129, 28130 数量:60台 出荷時期:平成30年9月〜令和7年3月 (2) X線透視診断装置 CoreVision LD 製造番号: 28006, 28007, 28009, 28010, 28011, 28012, 28015, 28017, 28020, 28022, 28025,...
Reason
Due to inadequate instructions regarding the replacement of the power plug, the plug was not replaced properly, resulting in smoke emission and the need to use an alternative medical device during surgery. Since smoke emission or a broken wire caused by a short circuit inside the power plug could render the device inoperable and affect the surgery, we are conducting a voluntary recall to replace the power plug.
Classification
Class II
Category
medical device recall
Country
Japan
Agency
PMDA-DEVICE
Status
Ongoing

Official recall notice from PMDA-DEVICE

Recall details on Nibble Recalls: https://trynibble.app/recalls/bcc26042-fd00-44a8-b86f-5e572a847a92