Heartsine Technologies - Defibrillator, Semi-automated

Recalling firm
Stryker New Zealand Ltd
Product
HeartSine Technologies - Defibrillator, semi-automated. [Device]. Model: SAM 350P and SAM 500P; SAM 360P. Affected: A subset of serial numbers beginning with 21, 22, 23 or 24, and are then followed by the letter B, D, E, G or H.; A subset of serial numbers beginning with 21, 22, 23 or 24, and are then followed by the letter B, D, E, G or H.. Also covers: HeartSine Technologies - Non-rechargeable public automated external defibrillator
Reason
23/7/26: Update - The manufaturer has identified a problem with the membrane tail, an internal component, which may result in an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. If this failure occurs, the device becomes inoperable., The manufacturer has identified that a limited number of HeartSine® Samaritan® PAD devices were distributed without the required maintenance supplement. This supplement contains periodic maintenance instructions implemented as an...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-07-23
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/bc56efa9-adaf-4c7e-8b9e-9b92e63d5196