Getinge Steam Sterilizer

Recalling firm
Getinge Australia Pty Ltd (NZ)
Product
Getinge Steam Sterilizer. [Device]. Model: HS 6610 AR1, HS 6613 ER-1, HS 6613 AR-2, HS 6613 AC-1, HS 6610 AC-1, HS 6610 EC-1, HS 6610 AR-1, HS 6613 AR-1; Getinge Steam Sterilizer Door. Affected: HS 5510 AC-1, HS 6613 DR-2, HS 5510 EC-1, HS 6610 ER-1, HS 6613T AR-1, HS 6613T AR-2, HS 6613 ER-2, HS 5510 ER-1. Also covers: LTSF Sterilizer
Reason
The manufacturer received a report of a user injury involving the autoclave door. The sterilization technician activated the door lock while a basket was obstructing the door's movement. Rather than reopening the door and repositioning the basket, the technician attempted to manually adjust the basket while the door was closing. When the basket became dislodged, the door closed and resulted in a serious crushing injury to the technician's hand., During further investigation, a second complaint involving a similar device was also identified.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-17
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE