Trilogy Evo
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Trilogy EVO. [Device]. Model: Trilogy Evo, Trilogy Evo O2. Affected: Update to the User Manual addendum
- Reason
- Analysis of the impact of nebulisers on performance of Trilogy Evo Platform ventilators revealed that using non-pneumatic nebulisers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient., Non-pneumatic nebulisers change the airflow characteristics by adding liquid droplets, causing the ventilator's leak estimation calculation to be incorrect. This leads to under-delivery of therapy to the patient even though the graphic user interface does not display the change. This condition may occur with or without...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-02-23
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE