Trilogy Evo

Recalling firm
Philips New Zealand Commercial Limited
Product
Trilogy EVO. [Device]. Model: Trilogy Evo, Trilogy Evo O2. Affected: Update to the User Manual addendum
Reason
Analysis of the impact of nebulisers on performance of Trilogy Evo Platform ventilators revealed that using non-pneumatic nebulisers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient., Non-pneumatic nebulisers change the airflow characteristics by adding liquid droplets, causing the ventilator's leak estimation calculation to be incorrect. This leads to under-delivery of therapy to the patient even though the graphic user interface does not display the change. This condition may occur with or without...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-23
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE