Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(ous Only); Software Version Number - All Versions: R1.x, R2.x, R3.x

Philips is recalling certain Azurion 7 M20 imaging systems because the patient table may move unexpectedly when the Reset Geometry button is pressed, even if the table lock is engaged. This affects systems with model numbers 722079, 722224, 722234, and 722282, running any R1.X, R2.X, or R3.X software versions. The recall applies to Azurion 7 M20 systems used in healthcare facilities.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
Product
Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Reason
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.