Philips North America: Philips Ingenia 1.5t Cx With Mr Elastography (mre). 1. Model Number (ref): 781262

Recall Summary Philips is recalling the Ingenia 1.5T CX MRI machine with MR Elastography (model 781262) because certain combinations of image reconstruction settings can produce inaccurate stiffness measurements during scans. The device is distributed worldwide, including throughout the United States and 50+ other countries. This is classified as a Class II recall, meaning the problem could cause incorrect medical results but is not immediately life-threatening.

Recalling firm
Philips North America
Product
Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-04-14
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...