Philips North America: Philips Ingenia 1.5t Cx With Mr Elastography (mre). 1. Model Number (ref): 781262
Recall Summary Philips is recalling the Ingenia 1.5T CX MRI machine with MR Elastography (model 781262) because certain combinations of image reconstruction settings can produce inaccurate stiffness measurements during scans. The device is distributed worldwide, including throughout the United States and 50+ other countries. This is classified as a Class II recall, meaning the problem could cause incorrect medical results but is not immediately life-threatening.
- Recalling firm
- Philips North America
- Product
- Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.
- Reason
- The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-04-14
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...
Official recall notice from U.S. Food and Drug Administration
Recall details on Nibble Recalls: https://trynibble.app/recalls/bc075fae-67b3-42d8-a3cf-412573bbe00b