Resectoscope Resection Sheaths

Recalling firm
Olympus (NZ) Ltd
Product
Resectoscope Resection Sheaths. [Device]. Model: RESECTION SHEATH and INNER SHEATH. Affected: Multiple lots
Reason
The manufacturer has identified complaints related to the ceramic tip of the resection/inner sheath breaking. There were reports, where the tip fell off during the procedure. In most events, the clinician identified and retrieved the fragment during the procedure. By nature, ceramics are not pliable and may fracture unexpectedly under significant mechanical impact. The manufacturer' investigation revealed that some devices remain in use for prolonged periods and may be subjected to stress over their operational lifetime., The manufacturer is highlighting the importance of strictly adhering...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/ba6d8521-c5aa-471d-8fc6-84be08c59f8b