Accuray Incorporated: Cyberknife Treatment Deliver System Equipped With The Xchange Robotic Collimator Changer; Linear Accelerator

Recall Summary Accuray's Cyberknife Treatment Delivery System with Xchange Robotic Collimator Changer is being recalled because a collimator (a radiation-shaping device) may not fully dock properly, and if room personnel then activate motion using a control pendant in the treatment room, the collimator could release and fall from the machine. The recall affects systems distributed worldwide, including throughout the United States and multiple other countries.

Recalling firm
Accuray Incorporated
Product
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Reason
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-06-16
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico,...