Tobii Dynavox Llc: System, Communication, Powered
Recall Summary Tobii Dynavox is recalling the TD Navio Midi Speech Generating Device (Model 1000798) because the battery mounting tabs can become damaged, causing the battery to shift inside the device and potentially create an internal short circuit that generates heat and releases gas. The device was distributed nationwide across all U.S. states and territories. This is a Class II recall, indicating a potential serious health hazard.
- Recalling firm
- Tobii Dynavox Llc
- Product
- TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
- Reason
- Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-07-08
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- US Nationwide distribution.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/b79bb99e-81f9-4f7c-bc7b-6936368a105a