Philips North America: Philips Mr 7700 With Mr Elastography (mre). 1. Model Number (ref): 782120. 2. Model Number (ref): 782153
Recall Summary Philips is recalling two models of MR 7700 MRI machines (model numbers 782120 and 782153) due to potential errors in stiffness measurements when certain image reconstruction settings are used. The error occurs with Resoundant's algorithm and can cause the voxel size settings in the default scan protocol to display incorrectly. The recall affects hospitals and medical facilities worldwide, including all U.S. states.
- Recalling firm
- Philips North America
- Product
- Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.
- Reason
- The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-04-14
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...
Official recall notice from U.S. Food and Drug Administration
Recall details on Nibble Recalls: https://trynibble.app/recalls/b5a8b90e-6991-44a9-8257-ccfa4384f4d9