Getinge Assured Batch Monitoring Pcd

Recalling firm
Getinge Australia Pty Ltd (NZ)
Product
Getinge Assured Batch Monitoring PCD. [Device]. Model: Model number - PL589/ 6005500575. Affected: 54043A
Reason
This problem was initiated by a customer product complaint regarding a missing O-ring. There have been no reported testing failures or adverse events, and the defect is unlikely to cause harm to the user. The root cause was identified as a combination of a supplier production process failure and an internal quality inspection error., Failure of the device, due to the missing O-ring, may state sterilisation cycle was not completed and may require re-sterilisation, which may disrupt workflow. This could result in delay of goods and, as such, delay of procedure; should this occur, it is...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-17
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE