Ssy,usm,is4000, Da Vinci Xi Surgical System, Da Vinci X Surgical System

Recalling firm
Device Technologies New Zealand
Product
SSY,USM,IS4000, da Vinci Xi Surgical System, da Vinci X Surgical System. [Device]. Model: ASSY,USM,IS4000, da Vinci Xi Surgical System, da Vinci X Surgical System. Affected: Product Codes: 380647
Reason
The manufacturer has identified that the screws on the cannula mount of the USM are from a specific population that are susceptible to breaking over time. When the cannula mount experiences a breakage of all six screws, the cannula mount can separate from the rest of the USM. Breakage of any fewer than six screws will not result in cannula mount separation, or impact to clinical performance.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-15
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/b3c9d6f1-5d01-4a5c-881b-1b284e205238