Medline Industries, Lp: Hysterectomy Kit
Recall Summary Medline Industries is recalling hysterectomy kits that contain CONMED AirSeal Access Ports due to a defect that can cause false pressure alerts and air leaks during surgery. These faulty ports may force surgeons to delay or stop procedures. The recall affects two specific kit models distributed nationwide, with particular concentration in Arizona, California, Indiana, Maryland, Missouri, New Jersey, Ohio, and Wisconsin.
- Recalling firm
- Medline Industries, LP
- Product
- Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135A CHN GYN ONCOLOGY DYNJ906731M
- Reason
- Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-08-31
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.