Staar Surgical Company Stella 2.0 Implant Orientation Diagram (iod)

Recalling firm
Device Technologies New Zealand
Product
STAAR Surgical Company Stella 2.0 Implant Orientation Diagram (IOD). [Device]. Model: MYOPIC TORIC EVO VTICMO12.6-10.5. Affected: Serial number - T2717072
Reason
The manufacturer has become aware that a limited number of IODs generated have the risk of an IOD with an incorrect axis value that could be clinically significant. Specifically, per the review of data in Stella 2.0, the IOD generated for a device you had ordered has a different axis placement than what had been reserved in Stella 2.0., Investigation points to an issue arising from migration of data from Stella 1.0 to Stella 2.0 associated with Made to Order (MTO) Toric ICLs. This data migration occurred prior to release of Stella 2.0 on April 13, 2026. The following three failure modes...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-17
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/afeaf3fe-885f-4563-8bf6-0433372f07ca