Uhi-4_high Flow Insufflator Unit

Recalling firm
Olympus (NZ) Ltd
Product
UHI-4_High Flow Insufflator Unit. [Device]. Model: UHI-4 High Flow Insufflation Unit. Affected: All afected Lot numbers
Reason
The manufacturer identified an increase in UHI-4 front-panel blackout complaints, where the front-panel display turns off unexpectedly, and air delivery (insufflation) ceases. An investigation found these blackouts (fault code E03) were primarily caused by the pressure sensors drifting positively from zero over time; when the pressure difference between the two sensors exceeds 7 mmHg, the UHI-4 generates the E03 fault, the front panel shuts off, and insufflation stops., is working to implement a long-term solution for front panel blackout. In the interim, the UHI-4 pressure sensors must...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-09
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE