Uhi-4_high Flow Insufflator Unit
- Recalling firm
- Olympus (NZ) Ltd
- Product
- UHI-4_High Flow Insufflator Unit. [Device]. Model: UHI-4 High Flow Insufflation Unit. Affected: All afected Lot numbers
- Reason
- The manufacturer identified an increase in UHI-4 front-panel blackout complaints, where the front-panel display turns off unexpectedly, and air delivery (insufflation) ceases. An investigation found these blackouts (fault code E03) were primarily caused by the pressure sensors drifting positively from zero over time; when the pressure difference between the two sensors exceeds 7 mmHg, the UHI-4 generates the E03 fault, the front panel shuts off, and insufflation stops., is working to implement a long-term solution for front panel blackout. In the interim, the UHI-4 pressure sensors must...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-06-09
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/af365134-4d88-4ffe-a9be-6e59bfb89123