Howmedica Osteonics Corp.: Tray, Surgical, Instrument

Recall Summary Howmedica Osteonics Corp. is recalling TS Femoral Prep Trays (Catalog Number 6543-6-507) used in orthopedic surgery because the sterilization instructions may result in moisture remaining on the trays after sterilization. The trays were distributed to hospitals and surgical facilities in the U.S. and internationally, with 158 units sent outside the U.S.

Recalling firm
Howmedica Osteonics Corp.
Product
TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.
Reason
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-09-16
Status
Open, Classified
Distribution
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

Official recall notice from FDA-DEVICES