Medline Industries, Lp: Cardiac Catheterization Kit

Recall Summary Medline Convenience cardiac catheterization kits containing Swan-Ganz catheters are being recalled because the catheters have a defective blue hub on the proximal injectate lumen that can leak or break during use. This defect can cause leakage, lumen damage, breakage, infection, medication loss, and blood loss. The affected kits (model DYNJ68530A) were distributed nationwide.

Recalling firm
Medline Industries, LP
Product
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A
Reason
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-28
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.