Staar Surgical Ag - Evo Visian Icl V4c (vicmo) Standard
- Recalling firm
- Device Technologies New Zealand
- Product
- STAAR Surgical AG - EVO Visian ICL V4c (VICMO) standard. [Device]. Model: Visian EVO ICL V4c (VICMO) standard; Visian EVO ICL V4c (VTICMO) toric; Visian EVO+ (VICM5) standard; Visian EVO+ (VTICM5) toric; Visian ICL (VICH); Visian Toric ICL (VTICH). Also covers: STAAR Surgical AG - EVO Visian ICL V4c (VTICMO) toric; STAAR Surgical AG - EVO Visian + (VICM5) standard; STAAR Surgical AG - EVO+ Visian ICL (VTICM5) toric; STAAR Surgical - VICH lens; STAAR Surgical - VTICH lens
- Reason
- The manufacturer has become aware of problems that may affect information displayed on the Implant Orientation Diagram (IOD) that is generated in Stella 2.0. The IOD is used for the implantation of Toric EVO/EVO+ Implantable Collamer Lenses (ICLs)., The identified problems are as follows:, 1. Expected refraction, spherical equivalent, and/or model/version values may be incorrect on the generated IOD., 2. If the sphere value for a Toric lens reservation is null, Stella 2.0 will default to indicate a rotation value of 90° on the IOD., 3. Users' local office network firewall settings may...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-06-12
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE