Stryker Drill Bits
- Recalling firm
- Stryker New Zealand Ltd
- Product
- Stryker Drill Bits. [Device]. Model: IMN Instruments Locking Drill D4.2x360mm. Affected: Catalog# 2351-4236S - Lot numbers: K1534DF, K1534E0, K1534E5
- Reason
- The manufacturer has identified that specific lots of IMN Instruments Locking Drills, Catalog# 2351-4236S, were manufactured with an oversized flat at the coupling piece of the drill bit. This problem prevents the insertion of the drill into the coupling attachment. The problem was discovered via product complaints.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-07-20
- Status
- Ongoing
- Distribution
- New Zealand - Wholesale
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/abeb18ab-0e29-402a-a651-934c3e5f4dbf