Ge Medical Systems Information Technologies Inc: Monitor, Physiological, Patient (without Arrhythmia Detection Or Alarms)
Recall Summary GE's Portrait Core Services Software version 1.0.5 is being recalled because a software issue can cause patient monitoring to stop working at the Hub and Central Viewer. The update can also reset all alarm settings back to factory defaults, overriding any custom alarm configurations that were set up for individual patients or care areas. The software was distributed in Michigan and New York, as well as in several international locations.
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Product
- PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353
- Reason
- Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-07-20
- Status
- Open, Classified
- Areas affected
- MI, NY
- Distribution
- US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/abc831be-4a51-4f3e-81cc-6719e2374215