Nav- Passive Spheres
NAV Passive Spheres Recall Summary Medtronic is recalling certain lots of NDI Passive Sphere optical tracking devices and Medtronic Biopsy Needle Kits used during surgery because the reflective coating on some spheres may not work properly. The reflective material on affected devices may not reflect infrared light as intended, which could cause surgical navigation systems to lose tracking of instruments during procedures or provide inaccurate positioning information. The defect is not visible to the naked eye, so affected devices cannot be identified through visual inspection alone.
- Recalling firm
- Medtronic New Zealand Limited
- Product
- NAV- Passive Spheres. [Device]. Model: NDI Passive Sphere – Model No. 8801071, 8801074, 8801075 (specific lots); Medtronic Biopsy Needle Kit – Model No. 9733068 (specific lots). Also covers: CST - NAV - Surgical Procedure Kit, general purpose
- Reason
- The manufacturer c has identified variability in the reflective material applied to a limited number of NDI Sterile Passive Sphere lots. The reflective coating is used to reflect infrared light from optical tracking cameras, allowing optical navigation systems to detect and track instruments during surgical procedures., The affected spheres may not be detected or tracked as intended by compatible navigation systems. This may result in instrument verification failures, intermittent or complete loss of tracking, tracking status errors, or reduced navigation accuracy. The variation in...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-09-01
- Status
- Ongoing
- Distribution
- New Zealand - Healthcare Professional
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/a86f2fd6-35aa-4eb3-bfe1-2bf934320282