HD And Pd Disposables 35000 Iia Zurich
- Recalling firm
- Vantive Limited
- Product
- HD and PD disposables 35000 IIa Zurich. [Device]. Model: Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp), Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp – Extra Short), Transfer Set (MiniCap Extended Life PD Transfer Set with Twist Clamp – Extra Long). Affected: Product code - R5C4482, R5C4483, R5C4484
- Reason
- 9/19/26: Update - The manufacturer has provided additional guidance and recommends continuing to use peroxide-cured silicone tubing sets. For patients under 6 months of age (approximately 3.5 kg), shorter transfer sets should be prioritised where platinum-cured silicone tubing or other alternatives are unavailable., Issue 1:, An increase in complaints has been reported involving separation of the female connector from the transfer set during connection or disconnection. The sterile fluid path remains intact, with no resulting risk of microbial contamination., Issue, 2:, The MiniCap...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-10-09
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE