Remel Europe Ltd.: Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Recall Summary Remel Europe Ltd. is recalling Oxoid Agglutinating Sera, Salmonella 9-0 (lot R30957301) because affected batches are showing cross-reactions with a different Salmonella type (2-0) within 60 seconds, which could cause laboratory technicians to misidentify one Salmonella strain as another. This product is a laboratory testing device distributed nationwide across the U.S. The recall is classified as Class II, indicating the product could cause incorrect test results but poses no direct harm to consumers.
- Recalling firm
- Remel Europe Ltd.
- Product
- Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
- Reason
- Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-03-25
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA.