Intra-aortic Balloon Balloon Pump

Recalling firm
Getinge Australia Pty Ltd (NZ)
Product
Intra-Aortic Balloon Balloon pump. [Device]. Model: Product: Intra-Aortic Balloon CS100 IABP, CS300 IABP
Reason
The manufacturer has become aware of instances where the device did not provide the expected runtime during some portable IABP uses CS100/CS300 devices. The manufacturer recently completed additional testing on the CS100/CS300 batteries and identified the battery runtime and cycle specifications contained in the devices' Instructions for Use (IFU) were incorrect and need to be updated. No other CS100/CS300 IABP performance deficiencies were identified.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-07-28
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/a5fb06cb-fff3-48b1-9e70-b442343f96e1