Neuroworks/sleepworks

Recalling firm
Natus Neuro New Zealand Limited
Product
NeuroWorks/SleepWorks. [Device]. Model: Part: 041906 NeuroWorks / SleepWorks Software and PK1288 (USB) PK/EXP1290 (NET) SD PLUS FLEXI Amplifier. Affected: NeuroWorks / SleepWorks 10.0.0 and with SD PLUS FLEXI
Reason
The manufacturer has become aware of complaints related to display differences when monitoring live recordings. A complaint was received where the live EEG shown on a remote monitoring station differed from the live EEG on the acquisition computer and from the same study was subsequently reviewed. This affects only NeuroWorks / SleepWorks 10.0.0 used with the SD PLUS FLEXI amplifier.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-18
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/a5239a7b-0d27-437f-b1cb-034ace5c020a