Dxc 500i Clinical Analyzer

Recalling firm
Beckman Coulter NZ Ltd
Product
DxC 500i Clinical Analyzer. [Device]. Model: With ISE DxC 500 AU Module w/ISE, DxC 500i and Without ISE DxC 500 AU Module, DxC 500i. Affected: Software - 1.3.5 or earlier
Reason
The manufacturer has identified a software issue affecting DxC 500i Clinical Analyzer when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyser to revert the test units to the system default setting without notifying the user., The system does not provide a warning or notification to alert the user that the configured units have reverted to the default units.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-02
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/a4838bfa-3ab3-4efe-97a3-2e10f5e0ac3b