Soltive System

Recalling firm
Olympus (NZ) Ltd
Product
Soltive System. [Device]. Model: Olympus SOLTIVE SuperPulsed Laser System (SOLTIVE Laser) products, models Pro (TFL-SLS) and Premium (TFL-PLS). Affected: Models/SKU - TFL-SLS/ R5000769, TFL-PLS/ R5000768, TFL-CPLU, TFL-CSLU
Reason
The manufacturer has received complaints describing an abrupt loss of power and functionality of the SOLTIVE SuperPulsed Laser System, or an inability to power on the unit. The investigation determined that a potential defect in the 24V power supply module in some SOLTIVE Laser units may cause the system to become inoperable. In some cases smoke or a burning smell may also occur. By design, the issue causing any smoke or burning smell that is detected would be contained within the internal laser console enclosure and would be self-extinguishing.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-26
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/a40c36ba-9f15-48c8-a722-c8ed259ac4e8