Ge Medical Systems Information Technologies Inc: Computer, Diagnostic, Programmable

GE CASE v7.0 Cardiac Testing System Recall Summary GE Medical Systems is recalling the CASE v7.0 Cardiac Testing System (Model 3031234-001) worldwide because the printer's electrostatic discharge brush can become dislodged, causing electrical sparking at the printer head that may ignite thermal paper and cause burns. The system has been distributed across the United States, Canada, and numerous countries in Europe, Asia, Africa, and the Middle East. This is a Class II recall affecting diagnostic cardiac testing equipment used in medical facilities.

Recalling firm
GE Medical Systems Information Technologies Inc
Product
CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3
Reason
Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-24
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Bangladesh, Cambodia, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, South Korea, Lebanon, Libya, Lithuania, Malaysia, Myanmar, Netherlands, New...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/a17f3cba-a459-4be5-a239-529e92e469eb